UKAS Calibration.

Accredited calibration. Trustworthy procedures. Traceable data.

As a UKAS accredited laboratory (4296) you can be sure that our staff are competent and fully-trained, our instruments are calibrated and properly maintained, and our work is carried out in the correct conditions.

 

UKAS is an internationally recognised accreditation body and UKAS accreditation ensures full metrological traceability of data. We calibrate all makes and models of pipettes and repeat dispensers, manual and electronic.

 

 

Use our MAIL-IN SERVICE with a quick turnaround, or we can come to you for ON-SITE SERVICING.

All pipettes for servicing must be accompanied with a completed and signed UKAS Decontamination Declaration!

As Found UKAS calibration.

As Found calibrations provide a check of the current performance of your pipette, enabling you to validate previous work. No adjustment or servicing is performed. As Found is purely a statement of how the instrument is performing.

All levels include the following:

  • As-Found calibration carried out at manufacturer's, ISO8655 or your own specification (note your requirement on the Decontamination Declaration)
  • Calibration Date Sticker to show the next calibration due date (if defined on the Decontamination Declaration)
  • A Certificate of Calibration showing a pass or fail based on the target specification

To determine accuracy and precision, calibration is carried out at set volume points with a specified number of readings depending on the service level, as shown below:

10 readings 
at 3 volumes

Min, mid & max volumes

As Found UKAS
Art. Nos:

Single-channel
on-site: C8137-0000
mail-in: C8137-M000

8-channel
on-site: C8237-0000
mail-in: C8237-M000

12-channel
on-site: C8337-0000
mail-in: C8337-M000

 

10 readings 
at 2 volumes

Min & max
volumes

As Found UKAS
Art. Nos:

Single-channel
on-site: C8130-0000
mail-in: C8130-M000

8-channel
on-site: C8230-0000
mail-in: C8230-M000

12-channel
on-site: C8330-0000
mail-in: C8330-M000

 

10 readings 
at 1 volume

Fixed
volume

As Found UKAS
Art. Nos:

Single-channel
on-site: C8129-1000
mail-in: C8129-M000

5 readings 
at 3 volumes1)

Min, mid & max volumes

As Found UKAS
Art. Nos:

Single-channel
on-site: C8139-0000
mail-in: C8139-M000

8-channel
on-site: C8239-0000
mail-in: C8239-M000

12-channel
on-site: C8339-0000
mail-in: C8339-M000

5 readings 
at 2 volumes1)

Min & max
volumes

As Found UKAS
Art. Nos:

Single-channel
on-site: C8439-0000
mail-in: C8439-M000

8-channel
on-site: C8539-0000
mail-in: C8539-M000

12-channel
on-site: C8639-0000
mail-in: C8639-M000

5 readings 
at 1 volume1)

Fixed
volume

As Found UKAS
Art. Nos:

Single-channel
on-site: C8739-1000
mail-in: C8739-M000

1) IMPORTANT: By selecting 5 readings per volume it is with the understanding that it will NOT permit a conclusion on conformity with ISO8655. The minimum requirement for conformity is 10 readings.

If your pipette is not within the target specification, we can perform a Service and As Returned UKAS Calibration (if requested beforehand on the Decontamination Declaration).

Mail-in

Mail-in pretend.

With a Service & UKAS Calibration, adjustment and servicing is undertaken. The purpose is to confirm post-calibration accuracy. It does not determine if the pipette was providing correct measurements during its last period of use.

All levels include the following:

  • Pipette dismantled, cleaned internally, inspected, leak tested and, if necessary, worn parts replaced
  • Pipette externally cleaned
  • Calibration carried out at manufacturer's, ISO8655 or your own specification (note your requirement on the Decontamination Declaration)
  • If necessary, your pipette will be adjusted to comply and recalibrated to meet the required specifications
  • Certification of Calibration showing the results and any spare parts fitted
  • Calibration Date Sticker to show the next calibration due date (if defined on the Decontamination Declaration)

To determine accuracy and precision, calibration is carried out at set volume points with a specified number of readings depending on the service level, as shown below:

10 readings
at 3 volumes

Min, mid & max volumes

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8135-0000
mail-in: C8135-M000

8-channel
on-site: C8235-0000
mail-in: C8235-M000

12-channel
on-site: C8335-0000
mail-in: C8335-M000

10 readings
at 2 volumes

Min & max
volumes

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8132-0000
mail-in: C8132-M000

8-channel
on-site: C8232-0000
mail-in: C8232-M000

12-channel
on-site: C8332-0000
mail-in: C8332-M000

10 readings
at 1 volume

Fixed
volume

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8736-0100
mail-in: C8736-M100

5 readings 
at 3 volumes1)

Min, mid & max volumes

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8136-0000
mail-in: C8136-M000

8-channel
on-site: C8236-0000
mail-in: C8236-M000

12-channel
on-site: C8336-0000
mail-in: C8336-M000

5 readings 
at 2 volumes1)

Min & max
volumes

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8436-0000
mail-in: C8436-M000

8-channel
on-site: C8536-0000
mail-in: C8536-M000

12-channel
on-site: C8636-0000
mail-in: C8636-M000

5 readings 
at 1 volume1)

Fixed
volume

Service & UKAS Cal
Art. Nos:

Single-channel
on-site: C8736-0500
mail-in: C8736-M500

1) IMPORTANT: By selecting 5 readings per volume it is with the understanding that it will NOT permit a conclusion on conformity with ISO8655. The minimum requirement for conformity is 10 readings.

UKAS pre & post calibration.

Pre & Post Calibrations provide both a statement of how the instrument is performing "as-found", and then adjustment and servicing is undertaken to confirm "as-left" post-adjustment accuracy.

All levels include the following:

  • As-Found UKAS calibration carried out at manufacturer's, ISO8655 or your own specification (note your requirement on the Decontamination Declaration)
  • A Certificate of Calibration showing a pass or fail based on the target specifiction
  • Service. Pipette dismantled, cleaned internally, inspected, leak tested and, if necessary, worn parts replaced
  • Pipette externally cleaned (if stated on Decontamination Declaration and Purchase Order)
  • As-Left UKAS Calibration carried out at manufacturer's, ISO8655 or your own specification
  • If necessary, your pipette will be adjusted to comply and recalibrated to meet the required specifications
  • Certification of Calibration showing the results and any spare parts fitted
  • Calibration Date Sticker to show the next calibration due date (if defined on the Decontamination Declaration)

To determine accuracy and precision, calibration is carried out at set volume points with a specified number of readings depending on the service level, as shown below:

10 readings 
at 3 volumes

Min, mid & max volumes

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8135-0000
mail-in: C8135-M000

8-channel
on-site: C8235-0000
mail-in: C8235-M000

12-channel
on-site: C8335-0000
mail-in: C8335-M000

 

10 readings 
at 2 volumes

Min & max
volumes

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8132-0000
mail-in: C8132-M000

8-channel
on-site: C8232-0000
mail-in: C8232-M000

12-channel
on-site: C8332-0000
mail-in: C8332-M000

 

10 readings 
at 1 volume

Fixed
volume

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8736-0100
mail-in: C8736-M100

5 readings 
at 3 volumes1)

Min, mid & max volumes

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8136-0000
mail-in: C8136-M000

8-channel
on-site: C8236-0000
mail-in: C8236-M000

12-channel
on-site: C8336-0000
mail-in: C8336-M000

5 readings 
at 2 volumes1)

Min & max
volumes

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8436-0000
mail-in: C8436-M000

8-channel
on-site: C8536-0000
mail-in: C8536-M000

12-channel
on-site: C8636-0000
mail-in: C8636-M000

5 readings 
at 1 volume1)

Fixed
volume

UKAS pre & post
Art. Nos:

Single-channel
on-site: C8736-0500
mail-in: C8736-M500

1) IMPORTANT: By selecting 5 readings per volume it is with the understanding that it will NOT permit a conclusion on conformity with ISO8655. The minimum requirement for conformity is 10 readings.

UKAS certification with purchased pipettes.

Contact our servicing team:

 

Have a query regarding a pipette
that has already been mailed-in
to our laboratory for servicing?

 servicecentre@starlab.co.uk

Have a general servicing enquiry
or want to discuss a new or existing on-site clinic?

+44 1908 283810

 service@starlab.co.uk