UKAS Calibration

As a UKAS accredited laboratory (4296), you can be sure that our staff are competent and fully trained, our instruments are well maintained and calibrated, and all calibration is carried out in controlled conditions.

For UKAS calibrations, we use certified 5-place, 6-place, and SpeedCal multichannel balances. UKAS is an internationally recognised accreditation body, and UKAS accreditation ensures full metrological traceability of data.

As Found UKAS Calibration

As Found Calibration checks the performance of your pipette prior to servicing or as part of your regular checks. No adjustment of the pipette is made; this is purely a statement of how the pipette is performing: 

  • The pipette will be checked against the specified specification (manufacturer's, ISO 8655 or your own)
  • Calibration Sticker
  • Certificate of Calibration provided for each pipette
  • If your pipette is not within specification, we can perform a Service & UKAS Calibration (if requested beforehand)

Parts at extra cost for all levels.

 

Description 10 Readings,
at 3 Volumes
10 Readings
at 2 Volumes
10 Readings
at 1 Volume
5 Readings
at 3 Volumes1
5 Readings
at 2 Volumes1
5 Readings
at 1 Volume1
Volumes at which your pipette will be checked min., mid. & max volumes min. & max. volumes fixed volume min., mid. & max volumes min. & max. volumes fixed volume
Number of readings taken at each volume 10 readings 10 readings 10 readings 5 readings 5 readings 5 readings
Calibration sticker, including due date (if defined)
Certificate of Calibration for each pipette
If out of spec, Service & As Returned UKAS Calibration (if requested beforehand)
Ordering Information
Single-channel / Repeat dispenser C8137-0000 / C8137-M000 C8130-0000 / C8130-M000 C8129-0000 / C8129-M000 C8139-0000 / C8139-M000 C8439-0000 / C8439-M000 C8739-0500 / C8739-M500
8-channel C8237-0000 / C8237-M000 C8230-0000 / C8230-M000 n/a C8239-0000 / C8239-M000 C8539-0000 / C8539-M000 n/a
12-channel C8337-0000 / C8337-M000 C8330-0000 / C8330-M000 n/a C8339-0000 / C8339-M000 C8639-0000 / C8639-M000 n/a
1IMPORTANT: By selecting 5 readings per volume it is with the understanding that it will NOT permit a conclusion on conformity with ISO8655.

Service & UKAS Calibration

For pipettes requiring a service as well as a calibration to UKAS standards.

  • Pipette dismantled, cleaned internally, inspected, leak tested and if necessary worn parts replaced
  • Pipette externally cleaned
  • Calibration Sticker
  • If necessary, your pipette will be adjusted to comply and recalibrated to the specified specification (manufacture’s, ISO 8655 or your own)
  • Certificate of Calibration provided for each pipette

Parts at extra cost for all levels.

 

Description 10 Readings,
at 3 Volumes
10 Readings
at 2 Volumes
10 Readings
at 1 Volume
5 Readings
at 3 Volumes1
5 Readings
at 2 Volumes1
5 Readings
at 1 Volume1
Pipette dismantled, cleaned internally, inspected and leak tested
Worn parts replaced, if necessary2
Outside of pipette cleaned
Volumes at which pipette is checked min., mid. & max volumes min. & max. volumes fixed volume min., mid. & max volumes min. & max. volumes fixed volume
Number of readings at each volume 10 readings 10 readings 10 readings 5 readings 5 readings 5 readings
Adjustment to meet required spec (if necessary)
Calibration sticker, including due date (if defined)
Certificate of Calibration incl. spare parts fitted
Ordering Information
Single-channel / Repeat dispenser C8134-0000 / C8134-M000 C8131-0000 / C8131-M000 C8738-0100 / C8738-M100 C8138-0000 / C8138-M000 C8438-0000 / C8438-M000 C8738-0500 / C8738-M500
8-channel C8234-0000 / C8234-M000 C8231-0000 / C8231-M000 n/a C8238-0000 / C8238-M000 C8538-0000 / C8538-M000 n/a
12-channel C8334-0000 / C8334-M000 C8331-0000 / C8331-M000 n/a C8338-0000 / C8338-M000 C8638-0000 / C8638-M000 n/a
1IMPORTANT: 5 readings per volume will NOT permit a conclusion on conformity with ISO8655.
2 Worn parts, if required, will be replaced without prior notice up to max. £40 (or amount agreed on decontamination form).

Pre & Post

Comprises As Found Calibration to show how your pipette has been performing, prior to us performing a Service and final UKAS Calibration:

  • The pipette will be checked against the specified specification (manufacturer's, ISO 8655 or your own)
  • (As Found) Certificate of Calibration provided for each pipette
  • Pipette dismantled, cleaned internally, inspected, leak tested and if necessary worn parts replaced
  • Pipette externally cleaned
  • Calibration Sticker
  • If necessary, your pipette will be adjusted to comply and recalibrated to the specified specification (manufacture’s, ISO 8655 or your own)
  • Certificate of Calibration provided for each pipette

Parts at extra cost for all levels.

 

Description 10 Readings,
at 3 Volumes
10 Readings
at 2 Volumes
10 Readings
at 1 Volume
5 Readings
at 3 Volumes1
5 Readings
at 2 Volumes1
5 Readings
at 1 Volume1
AS FOUND CALIBRATION
Volume at which pipette is checked min., mid. & max volumes min. & max. volumes fixed volume min., mid. & max volumes min. & max. volumes fixed volume
Number of readings at each volume 10 readings 10 readings 10 readings 5 readings 5 readings 5 readings
Certificate of As Found Calibration
SERVICE
Pipette dismantled, cleaned internally, inspected and leak tested
Worn parts replaced, if necessary2
Outside of pipette cleaned
AS LEFT CALIBRATION
Volume at which pipette is checked min., mid. & max volumes min. & max. volumes fixed volume min., mid. & max volumes min. & max. volumes fixed volume
Number of readings at each volume 10 readings 10 readings 10 readings 5 readings 5 readings 5 readings
Adjustment to meet required specification (if needed)
Calibration sticker incl. due date (if defined)
Certificate of Calibration incl. spare parts fitted
Ordering Information
Single-channel / Repeat dispenser C8135-0000 / C8135-M000 C8132-0000 / C8132-M000 C8736-0100 / C8736-M100 C8136-0000 / C8136-M000 C8436-0000 / C8436-M000 C8736-0500 / C8736-M500
8-channel C8235-0000 / C8235-M000 C8232-0000 / C8232-M000 n/a C8236-0000 / C8236-M000 C8536-0000 / C8536-M000 n/a
12-channel C8335-0000 / C8335-M000 C8332-0000 / C8332-M000 n/a C8336-0000 / C8336-M000 C8636-0000 / C8636-M000 n/a
1IMPORTANT: 5 readings per volume does not allow ISO8655:6:2022 conformity determination (minimum requirement is 10 readings).
2 Worn parts will be replaced without prior contact up to max. value £40 (or amount agreed on decontamination form).

On Purchased Pipette

Do you need your new pipette supplied with UKAS Certification?

We can supply your new pipette from Starlab with a UKAS Certificate of Calibration. Simply order the calibration at the same time as ordering your new pipette. Starlab will perform a 3.10 As Found Calibration on the pipette prior to shipping, and enclose the Calibration Certificate with your pipette.

To order the calibration at the same time as your new pipette, see the Accessories tab for the pipette you wish to purchase, and add the calibration to your shopping basket.

 

Logo for pipete calibration of new pipettesLogo for pipete calibration of new pipettes

Contact Pipette service

  • For any query regarding a pipette that has been mailed-in to our laboratory, please contact servicecentre@starlab.co.uk
  • For general servicing enquiries, to book a clinic or if you have a query regarding an existing clinic, please contact service@starlab.co.uk or 01908 283810
Logo for contacting Starlab service teamLogo for contacting Starlab service team